< back to products page
Analyzer for detection of components in water
- Manufacturer - Waltron, USA
- Catalogue of some of the presented products
 µAI 9049 Ethylene Glycol Analyzer
REVOLUTIONARY PATENTED TECHNOLOGY
Our Colorimetric Ethylene Glycol Analyzer (Aqualyzer® Series 9040) is based on the new patented Loop Flow Analysis (LFA) cycle.
LFA technology involves isolating,manipulating and testing a known volume of sample in a closed loop system. A powerful programmable microprocessor allows for customized program cycles,ensuring accuracy and reliability with every application.
|
GENERAL SPECIFICATIONS / Performance |
Sensor Classification | Colormetric with glass flowcell and LED source |
Application | Demineralized,boiler,potable,surface and waste water |
Power Requirements | 12V |
Power Supply | 110-120V or 220-240V 50/60 Hz, 100 VA |
Humidity | Up to 90% not condensable |
Ambient Temperature Range | 10-45 °C analyzer (50-113° F) |
Range | Silica: 0-500 ppb, 0-5000 ppb, 0-200 ppm* Phosphate: 0-2.4 ppm,0-7 ppm, 0-20 ppm, 0-200 ppm* Hydrazine, Copper: 0-100 ppb, 0-1 ppm* Ammonia: 0-750 ppb, 0-7.5 ppm* Ethylene Glycol: 0-15 ppm, 0-150 ppm* *By Auto Dilutio |
Accuracy | Silica: +/- 1 ppb or +/- 2% of reading, whichever is greater Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/- 2% of full scale |
Unit Dimensions | Height = 31.5" (80.01 cm), Width = 16.54" (42.01 cm), Depth = 11.13" (28.27 cm) |
Atmospheric Pressure Range | No Limits |
Effect of Electromagnetic Fields | EMC tested according to CE compliance |
Tolerance to Electrostatic Discharges | EMC tested according to CE compliance |
Positioning and installation details | Wall or panel mounting analyzer,installed approx 39" (99cm) from the ground,maximum 13 (396cm) from sampling point |
Response Time (approximate) | Silica: 10 min, Phosphate: 6 min, Hydrazine: 12 min, Ammonia: 12 min, Copper: 9 min, Ethylene Glycol: 10 min |
Alarms | Concentration, cal error, out of service/sample |
Reproducibility | Silica:+/ 1 ppb or +/ 2% of reading, whichever is greater Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/ 2% of reading |
Degree of Protection | IP55 (NEMA 4): Wet section, IP65 (NEMA 4x): Electronics |
Required Maintenance | Once a month replenish reagents, once a year replace tubing |
Reagent Consumption (approximate) | Silica, Phosphate, Ammonia, Copper: 1 liter each per month Hydrazine, Ethylene Glycol: 4 liters each per month |
Reactor Volume | 10ml |
Materials in Contact with Sample | Glass, Silicone, Plexiglass, Stainless Steel AISI 316 |
CE Compliance | Meets low voltage and low electromagnetic compatibility directives |
Sample Conditioning Requirements | Filtering between 10 and 60 micron, depending on the matrix, needed only to avoid clogging. Internal disposable filters supplied (60 micron) |
|
AQUALYZER® 9001 SERIES
 Sodium Analyzer Aqualyzer® 9031
The Waltron Aqualyzer® 9031 Sodium Analyzer is the most advanced online instrument specifically developed for use in continuous measurement and determination of sodium ion concentration.
Common application points in a boiler system for online sodium measurement include:make-up water,condensate,boiler feedwater and saturated steam/main steam.
The Aqualyzer® 9031 sodium analyzer operates automatically and comes complete with true fully automatic calibration.
|
GENERAL SPECIFICATIONS / Performance |
Range | 0.01ppb - 10ppm |
Accuracy | ± 5% of reading or ± 0.1ppb (whichever is greater) within ± 5°C of calibration temperature |
Reproducibility | ± 5% of reading or ± 0.1ppb (whichever is greater) at constant temperature |
Response Time | 90% of 1-10ppb step: less than 4 minutes; 90% of 100-1ppb step: less than 6 minutes |
Temperature Compensation | Automatic |
pH Compensation | Alkaline vapor buffer: 0.88 SG (35% W/V) ammonia above 0.1 ppb sodium, 50% diethylamine solution below 0.1ppb sodium |
Single or 2-point Calibration | Manual and fully automatic calibration. Manual initiated by operator. Automatic with programmable frequency in days. No manual changes of standards during calibration |
Current Outputs | Two isolated 4-20mA current (analog) outputs |
Digital Outputs | Communication interface via RS-232 & RS-485 using MODBUS RTU protocol |
Alarms | Three voltage-free contacts, alarm points set from transmitter rated at 2A, 250VAC (non-inductive) |
Power | Wide range of input power supply 90VAC 250VAC |
CE Certification (on request) | Meets low voltage and low electromagnetic compatibility directives |
Operating Conditions |
Sample | Temperature: 41 - 131°F (5 - 55°C); Flow 150 - 400 ml/min |
Ambient Temperature | 32 - 131°F (0 - 55°C) |
Pressure | 5 - 30 psig (0.35 - 2.07 bar) |
Composition | Sample should be filtered to 60 microns, free of film forming compounds |
Mechanical |
Construction | Transmitter high strength ABS, IP65 (NEMA 4x) Wet section high strength ABS and sheet metal, all wetted parts stainless steel or PVC,IP55 (NEMA 4) |
Dimensions | Height = 31 (78.74 cm) including drain, Width = 22 (55.88 cm), Depth = 8.5 (21.59 cm) |
Mounting | Four key-style holes: Height = 0.86 (2.18 cm), Top Width = 0.281 (0.714 cm), Radius = 0.280 (0.711 cm) Horizontal: 14.7 (37.34 cm), Vertical: 11 (27.94 cm) |
Weight | 40 lbs. (18.14 kg) |
Wiring | Eight (8) conduit knockouts |
Piping | Sample Inlet: 1/4" (0.635 cm) OD Swagelokt®, Drain: 3/8" (0.953 cm) OD tube |
Distance | Up to 325 feet (100 meters) on request |
Finish | Corrosion resistant |
|
 Dissolved Oxygen Analyzer Aqualyzer® 9061
The Waltron Analyzer® 9061 Dissolved Oxygen Analyzer is a state-of-the-art microprocessor-based instrumentation package developed specifically for use in measuring low and high concentrations of dissolved oxygen. It is todays standard for all low range dissolved oxygen analyzers.
True Fully Automatic Calibration
Calibration is carried out automatically online and involves no manual operation.An automatic calibration check is performed at a predetermined frequency by the instrument; however, calibration can be initiated at any time manually if needed.
Common application points in a boiler system for online dissolved oxygen measurement include: deaerator, condensate, boiler feed water and economizer inlet.
|
GENERAL SPECIFICATIONS / Performance |
Range | 0-1000 ppb,0-20 ppm with automatic range selection. |
Accuracy | ± 1 ppb |
Stability | ± 5% of reading or ± 2 ppb per week, whichever is greater |
Response Time | 90% of a step change in less than three minutes |
Precision | ± 0.5 ppb |
Temperature Compensation | Automatic,reduces sample temperature error to less than 0.5% of reading per degree C |
Calibration | Automatic,reduces sample temperature error to less than 0.5% of reading per degree C |
Current Outputs | Two isolated 4-20mA current (analog) outputs. |
Digital Output | Communication interface via RS-232 and RS-485 using MODBUS RTU protocol |
Alarms | Three voltage-free contacts. Alarm points set from transmitter rated at 2A,250VAC (non-inductive |
Response Time | 90% of a step change in 60 second |
Atmospheric Pressure Compensation | Preset to normal atmospheric pressure at sea level and programmable |
Thermal Protection | Automatic sample bypass if sample exceeds 131°F (55°C) |
Power | 90VAC-250VAC, wide range of input power supply |
CE Certification (on request) | Meets low voltage and low electromagnetic compatibility directives |
Operating Conditions |
Sample Flow | 100 300 ml/min |
Sample Temperature | 41 - 131° F (5 - 55°C) |
Ambient Temperature | 32 - 131° F (0 - 55°C) |
Pressure | 530 psig (0.35 - 2.07 bar) |
Recommended Conditioning | Water sample filtering to 60 micron |
Mechanical |
Construction | Transmitter high strength ABS, IP65 (NEMA 4x) Wet section high strength ABS and sheet metal, all wetted parts stainless steel or PVC, IP55 (NEMA 4) |
Dimensions | Transmitter:Height = 10.92 (27.74 cm), Width = 7 (17.78 cm), Depth = 3.54 (8.99 cm) Wet Section: Height = 18.25 (46.36 cm) including drain, Width = 11.25 (28.58 cm), Depth = 7.63 (19.38 cm) |
Mounting | Four key-style holes: Height = 0.86 (2.18 cm), Top Width = 0.281 (0.714 cm), Radius = 0.28 (0.711 cm) Horizontal: 9 (22.86 cm), Vertical: 12.875 (32.70 cm) |
Weight | 23 lbs. (10.44 kg) |
Wiring | Eight (8) conduit knockouts |
Piping | Sample Inlet: 1/4" (0.635 cm) OD Swagelokt, Drain: 3/8" (0.953 cm) OD tube |
Distance | Up to 325 feet (100 meters) on request |
Finish | Corrosion resistant |
|
 Hydrazine Analyzer Aqualyzer® 9071
The Waltron Aqualyzer® 9071 Hydrazine Analyzer is the most advanced hydrazine instrument specifically developed for use in continuous measurement and determination of hydrazine concentration in boiler feedwater.
Common application points in a boiler system for online hydrazine measurement include:deaerator,condensate system and boiler feedwater.
The Aqualyzer® 9071 hydrazine analyzer operates automatically and comes complete with true fully automatic calibration.
|
GENERAL SPECIFICATIONS / Performance |
Range | 0 - 1000ppb (1ppm) |
Accuracy | ±5% of reading or ±2ppb, whichever is greater for concentrations up to 500ppb. 10% or better for concentrations above 500ppb |
Stability | ± 5% of reading or ± 2ppb per week, whichever is greater |
Response Time | 90% of a step change in less than 3 minutes |
Temperature Compensation | Automatic, reduces sample temperature error to less than 0.5% of reading per degree C |
pH Compensation | Buffer continuously induced into sample flow Ammonium Hydroxide 0.88 SG (35% WN) |
Calibration | Manual and fully automatic calibration. Automatic with programmable frequency in days. Manual initiated by operator.No manual changes of standard during calibration. |
Current Outputs | Three isolated 4-20mA current (analog) outputs |
Digital Outputs | Communication interface via RS-232 & RS-485 using MODBUS RTU protocol |
Alarms | Three voltage-free contacts,alarm points set from transmitter rated at 2A, 250VAC (non-inductive) |
Power | Wide range of input power supply 90VAC - 250VAC |
CE Certification (on request) | Meets low voltage and low electromagnetic compatibility directives |
Óñëîâèÿ çà ðàáîòà |
Sample | Temperature: 41 - 131° F (5 - 55° C); Flow 150 400 ml/min |
Ambient Temperature | 32 - 131° F (0 - 55° C) |
Pressure | 530 psig (0.35 - 2.07 bar) |
Composition | Sample should be filtered to 60 microns, free of film forming compounds |
Ìåõàíè÷íè |
Construction | Transmitter high strength ABS, IP65 (NEMA 4x) Wet section high strength ABS and sheet metal, all wetted parts stainless steel or PVC, IP55 (NEMA 4) |
Dimensions | Height = 31 (78.74 cm) including drain, Width = 22 (55.88 cm), Depth = 8.5 (21.59 cm) |
Mounting | Four key-style holes: Height = 0.86 (2.18 cm), Top Width = 0.281 (0.714 cm), Radius = 0.28 (0.711 cm) Horizontal: 14.7 (37.34 cm), Vertical: 11 (27.94 cm) |
Weight | 40 lbs. (18.14 kg) |
Wiring | Eight (8) conduit knockouts |
Piping | Sample inlet:1/4"(0.635 cm) OD Swagelokt®, Drain: 3/8" (0.953 cm) OD tube |
Distance | Up to 325 feet (100 meters) on request |
Finish | Corrosion resistant |
|
|